How does an Oxygen Concentrator Work?

An oxygen concentrator is a medical device that concentrates oxygen from ambient air. Atmospheric air has about 78 per cent nitrogen and 21 per cent oxygen, with other gases making up the remaining 1 per cent. The oxygen concentrator takes in this air, filters it through a sieve, releases the nitrogen back into the air, and works on the remaining oxygen.


This oxygen, compressed and dispensed through a cannula, is 90-95 per cent pure. A pressure valve in concentrators helps regulate supply, ranging from 1-10 litres per minute.


According to a 2015 report by the WHO, concentrators are designed for continuous operation and can produce oxygen 24 hours a day, 7 days a week, for up to 5 years or more.


Pressure swing adsorption

These oxygen concentrators use an atomic sifter to adsorb gases and work on the rule of fast tension swing adsorption of environmental nitrogen onto zeolite minerals at high strain. This kind of adsorption framework is accordingly practically a nitrogen scrubber passing on the other air gases to go through, leaving oxygen as the essential gas remaining. Public service announcement innovation is a dependable and conservative procedure for little to mid-scale oxygen age. Cryogenic division is more reasonable at higher volumes and outside conveyance by and large more appropriate for little volumes.


At high strain, the permeable zeolite adsorbs enormous amounts of nitrogen, on account of its huge surface region and substance attributes. The oxygen concentrator packs air and ignores it zeolite, making the zeolite adsorb the nitrogen from the air. It then, at that point, gathers the excess gas, which is for the most part oxygen, and the nitrogen desorbs from the zeolite under the diminished strain to be vented.


An oxygen concentrator has an air compressor, two chambers loaded up with zeolite pellets, a tension balancing repository, and a few valves and cylinders. In the principal half-cycle, the primary chamber gets air from the blower, which goes on around 3 seconds. During that time the tension in the main chamber ascends from barometrical to about 2.5 occasions ordinary climatic strain (regularly 20 psi/138 kPa check, or 2.36 airs outright) and the zeolite becomes soaked with nitrogen. As the primary chamber comes to approach unadulterated oxygen (there are limited quantities of argon, CO2, water fume, radon and other minor barometrical parts) in the principal half-cycle, a valve opens and the oxygen-improved gas streams to the strain leveling repository, which associates with the patient's oxygen hose. Toward the finish of the primary portion of the cycle, there is another valve position change with the goal that the air from the blower is coordinated to the subsequent chamber. The tension in the principal chamber drops as the advanced oxygen moves into the supply, permitting the nitrogen to be desorbed once more into gas. Mostly during that time half of the cycle, there is another valve position change to vent the gas in the primary chamber once again into the surrounding air, keeping the grouping of oxygen in the tension balancing repository from falling underneath about 90%. The tension in the hose conveying oxygen from the adjusting supply is kept consistent by a strain lessening valve.


Older units cycled for a period of about 20 seconds and supplied up to 5 litres per minute of 90+% oxygen. Since about 1999, units capable of supplying up to 10 L/min have been available.


Classic oxygen concentrators utilize two-bed sub-atomic sifters; more current concentrators use multi-bed sub-atomic strainers. The upside of the multi-bed innovation is the expanded accessibility and excess, as the 10 L/min atomic sifters are stunned and duplicated on a few stages. With this, more than 960 L/min can be created. The increase time - the slipped by time until a multi-bed concentrator is delivering oxygen at >90% focus - is frequently under 2 minutes, a lot quicker than basic two-bed concentrators. This is a major benefit in portable crises. The alternative, to fill standard oxygen chambers (for example 50 L at 200 bar = 10,000 L each) with high-pressure supporters, to guarantee programmed failover to recently filled hold chambers and to guarantee the oxygen production network for example if there should be an occurrence of force disappointment, is given with those frameworks.


Membrane separation

In membrane gas separation, membranes act as a permeable barrier which different compounds move across at different rates or do not cross at all.